Advanced Use - Fluke PM2811 User Manual

Programmable power supplies
Table of Contents

Advertisement

5 - 28
GPIB control:
GPIB control can only be done in the remote mode of operation. An external
trigger can be programmed in the following ways:
Send
TRG
*
Send
GET

5.6 ADVANCED USE

This section describes how to connect several output channels (with the same
ratings) to obtain higher output power through a connected load.
To increase the total output current through a connected load, the output channels
must be connected in parallel. Refer to section 5.6.1.
To increase the total output voltage over a connected load, the output channels
must be connected in series. Refer to section 5.6.2.
This section also describes the way to connect more than one load to an output
channel (multiple loads). Refer to section 5.6.3.
CAUTION: Only output channels with equivalent voltage and current ratings may
be connected in series or in parallel to prevent that the channel with
the lowest output ratings will be damaged.
By sending the common command
supply.
By sending the GET (Group Execute Trigger) code over
the GPIB interface.
Users Manual
TRG to the power
*