Getting Familair with the Power Supply
9)
Disable the overcurrent protection and reset the output channel by pressing:
OCP EN
OCP DIS
Check that the display reads approximately 0.5A.
Also check that the CC annunciator is on, indicating that the output channel
is in the 'Constant Current' mode again.
10) Turn off the power supply.
11) Remove the jumper wire between the - V(oltage) and + V(oltage) output
terminals and reattach the terminal block cover.
RESET
4 - 11