Installation; Mechanical Installation - Fluke PM2811 User Manual

Programmable power supplies
Table of Contents

Advertisement

3 - 2
The instrument described in this manual is to be used by properly trained
personnel.
Do not remove the cover or perform any adjustment, maintenance, or repair
unless you are qualified to do so and are aware of the hazards involved.
Before connecting the instrument to line power, ensure that the
Symbol
power ground is functioning correctly. Before any other connection
is made, the instrument must be connected to the earth (ground)
conductor via the three-conductor power cord. The power plug
must be inserted into a socket outlet provided with ground contact.
The grounding protection must not be defeated by use of an
extension cord without a ground conductor.
WARNING : Any interruption of the protective conductor inside or outside
the instrument or disconnection of the protective ground
terminal is likely to make the instrument dangerous.
Intentional interruption is prohibited.
The instrument is shipped from the factory with a power cord that has a plug
appropriate for the country in which it is sold. If a different power plug or power
cord is required, contact your Fluke Sales and Service Center.
CAUTION: When an instrument is brought from a cold into a warm environment,
condensation may cause a hazardous condition; therefore, make
sure that the grounding requirements are strictly adhered to.

3.4 INSTALLATION

3.4.1

Mechanical installation

The instrument is fan cooled and must be installed in a location that allows
sufficient space at the three important sides: rear, left, and right. Each side must
have a space of at least 1 inch (25 mm). The fan, located at the rear, cools the
instrument by drawing air into the openings at the rear and exhausting it through
openings on the sides. Each instrument output channel has its own fan.
Users Manual

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pm2812Pm2813Pm2831Pm2832