Fluke PM2811 User Manual page 77

Programmable power supplies
Table of Contents

Advertisement

Using your Programmable Power Supply
Remote control (automatic stepping):
The actual memory address index (1 to 999) can be selected using the
[:SOURce]:LIST:INDex[:NSELect] command. The selected memory address can
be requested by means of the [:SOURce]:LIST:INDex[:NSELect]? query.
A sequence list of memory addresses can be defined using the
[:SOURce]:LIST:SEQuence:STARt and [:SOURce]:LIST:SEQuence:STOP
commands.
A sequence list of memory addresses can be reported using the
[:SOURce]:LIST:SEQuence:STARt? and [:SOURce]:LIST:SEQuence:STOP?
queries.
The length of a sequence list, i.e., the number of voltage and current points, can
be requested by the [:SOURce]:LIST:VOLTage:POINts? and
[:SOURce]:LIST:CURRent:POINts? queries.
The step interval time can be defined by the [:SOURce]:LIST:TIMer command and
can be requested by the [:SOURce]:LIST:TIMer? query.
The source of the next step in the sequence list (the step trigger) can be defined
and requested by the [:SOURce]:LIST:SOURce command, respectively the
[:SOURce]:LIST:SOURce? query. The following sources are possible:
If source = TIMer, the internal timer of the power supply is used to count
the step interval time.
If source = EXTernal or BUS, the active trigger source depends on whether
the START line of the trigger bus at the rear of the power supply is active
or not. If active, step pulses from the external STEP input will cause the
next step to be performed. If not active, the next step will be performed
after the
*
The execution of a sequence list can be started using the :INITiate[:IMMediate]
command. The first address is the START address, and the last address is the
STOP address.
A sequence list can be executed continuously by means of the
:INITiate:CONTinuous command.
The execution of a sequence list can be stopped by means of the :ABORt
command. If the execution is stopped, the last recalled voltage and current setting
remain active.
TRG command or after the GET trigger command.