Table of Contents Section 1: Introduction....................1-1 General Description ..................1-1 Application ....................1-1 Specifications....................1-1 Receiving Inspection and Storage ............... 1-2 Installation....................1-3 Procedures, Warnings, and Cautions ............1-3 Section 2: Operation....................2-1 960PS-200 Power Supply................2-1 Input Filter....................2-1 DC to DC Converter..................
Introduction General Description Section 1 Introduction 1.1 General Description The 960PS Power Supply is used to supply power to the modules that are plugged into the 960MB/961MB motherboards. The power supply is mounted outside of the 960CC Card Cage Module, the wall mounted Local Control Unit enclosure, or control room cabinet.
Upon receipt of the unit: 1. Inspect the carton(s) and contents for damage. If damage is evident, file a claim with the carrier and notify Fluke Biomedical, Radiation Management Services at 440.248.9300. 2. Remove the contents from the packing material.
Biomedical. It shall be the owner's or user's responsibility to see to it that the procedures described here are meticulously followed, and especially that WARNINGS and CAUTIONS are heeded. Failure on the part of the owner or user in any way to follow the prescribed procedures shall absolve Fluke Biomedical and its agents from any resulting liability.
Operation 960PS-200 Power Supply Section 2 Operation 2.1 960PS-200 Power Supply The 960PS-200 supplies three different DC voltage outputs to the 960 modules through the 960MB/961MB motherboard. 115 VAC is connected to terminal strip A3, part of the front panel. The AC line voltage is fed to a bridge rectifier circuit and rectified diodes CR1, CR2, CR3, and CR4.
Victoreen 960PS-200 & 960PS-210 Operators Manual 2.5 960PS-210 Power Supply The 960PS-210 Power Supply is similar in operation to the 960PS-200, with the exception that it incorporates an additional 24 VDC output, an input voltage sensing circuit, and a common and differential mode choke.
There are no user adjustments on the main 960PS-200. Should the output voltage fall outside of the tolerance specified, the unit must be returned to Fluke Biomedical for repair. Adjustments to the 960PS-210 are limited to the AC Line Monitor Voltage trip point adjustment (R40), mounted on the stand-up circuit board, and the +5 VDC output voltage adjustment (R17).
Page 12
Victoreen 960PS-200 & 960PS-210 Operators Manual If a problem develops, verify that the voltages at connection point inputs and outputs are present and that all wiring is secure. Ensure fuse F1 is not open for the 960PS-210 Power Supply. Refer to Appendix B for drawings.
Appendix Connector Designations Appendix A Connector Designations A.1 Connector Designations Terminal Designations, 960PS-200 Description AC Line Voltage AC Line Voltage -15 @ 2 A +15 @ 2 A RETURN +5 V Return Sense +5 V@ 10 A + 5 V SENSE CHASSIS GND Terminal Designations, 960PS-21 0 Description...
Schematic Diagram B.2 Applicable Bill of Materials The 960PS-200 Power Supply is manufactured specifically for Fluke Biomedical by Adtec Power, Inc. and the 960PS-210 Power Supply is manufactured specifically for Fluke Biomedical by Deltron, Inc. These devices are not user serviceable, and sub-component bill of materials is not available. The unit must be...