5 - 20
At delivery the contents of all addresses are set to their minimum value. When an
address has been filled, its contents can be recalled, so that the settings become
actual. The actual settings can be updated in OPERATE and STANDBY modes,
but they are only active in the OPERATE mode. When the power supply is in the
STANDBY mode, the actual settings will become active as soon as the OPERATE
mode is entered.
Note:
To remember which locations have been programmed and which have
not, keep a record (such as the following) of which locations have been
used.
MEMORY
ADDRESS
01
02
03
04
Local control (store/recall):
The voltage V
set
The output voltage can be adjusted using the
The current I
can be set using the
set
The output current can be adjusted using the
The actual voltage/current setting can be stored using the
The actual voltage/current setting can be recalled using the
Example:
REQUIREMENTS : Fill the memory address 23 with 5V/0.3A and the
SOLUTION:
OPR
Press
STBY
For multiple output units, press
channel 1 has been selected (see indication 1 on the display).
Press
V
Press
I
Press
STORE
OUTPUT CHANNEL 1
VOLTAGE
2.0
1.9
1.8
etc.
can be set using the
memory addresses 33, 34, and 35 with, respectively,
4V/1A, 3V/1A and 2V/1A for output channel 1.
to make the STANDBY mode active.
to set the actual voltage at 5V.
5
ENTER
3
ENTER
to set the actual current at 0.3A.
to store the 5V and 0.3A at memory address 23.
2
3
ENTER
CURRENT
VOLTAGE
0.00
0.10
0.20
etc.
V
key and the numeric input keys.
V
key and the numeric input keys.
I
I
a number of times until output
SELECT
Users Manual
OUTPUT CHANNEL 2
CURRENT
5.9
5.8
5.7
etc.
key in the OPERATE mode.
key in the OPERATE mode.
key.
STORE
RECALL
key.
0.90
0.80
0.70
etc.