Chapter 14, Oscilloscope Screen
Field Descriptions
Oscilloscope Trigger Menu Fields
Figure 85
Oscilloscope Triggering
Internal
Auto/Norm
396
This field selects the trigger source.
•
Internal uses the signal being displayed for triggering.
•
External uses the rear-panel EXT SCOPE TRIGGER INPUT for triggering. This is
a TTL level trigger (approximately 2.5 V).
1
uses the signaling encoder for triggering. The encoder must be sending its
•
ENCODER
signal to trigger the oscilloscope.
This field specifies how the trigger level is set.
•
Auto automatically triggers if a triggering signal is not detected within approximately
50 ms of the last trigger.
•
Norm requires a specific triggering signal before triggering.
Operating Considerations
Automatic triggering should be used for signals >20 Hz. Normal triggering should
be used for signals 20 Hz.
Also, when measuring 1 Hz signals, you should set the
ANALYZER screen to
1.
Optional on some Test Set models
to provide dc coupling to the oscilloscope's input.
Input
oscscrn4.wmf
field in the AF
Scope To