Chapter 14, Oscilloscope Screen
Field Descriptions

Oscilloscope Trigger Menu Fields

Figure 85
Oscilloscope Triggering
Internal
Auto/Norm
396
This field selects the trigger source.
Internal uses the signal being displayed for triggering.
External uses the rear-panel EXT SCOPE TRIGGER INPUT for triggering. This is
a TTL level trigger (approximately 2.5 V).
1
uses the signaling encoder for triggering. The encoder must be sending its
ENCODER
signal to trigger the oscilloscope.
This field specifies how the trigger level is set.
Auto automatically triggers if a triggering signal is not detected within approximately
50 ms of the last trigger.
Norm requires a specific triggering signal before triggering.
Operating Considerations
Automatic triggering should be used for signals >20 Hz. Normal triggering should
be used for signals 20 Hz.
Also, when measuring 1 Hz signals, you should set the
ANALYZER screen to
1.
Optional on some Test Set models
to provide dc coupling to the oscilloscope's input.
Input
oscscrn4.wmf
field in the AF
Scope To