HP 8921A User Manual page 293

Cell site test set
Hide thumbs Also See for 8921A:
Table of Contents

Advertisement

Detector
This setting selects the type of detector used when measuring and displaying AF
signal levels.
Detector Types
RMS displays the RMS value of signals.
displays the RMS value of a signal multiplied by the square-root-of-2.
Pk+ displays the positive peak value.
Pk displays the negative peak value.
Pk /2 adds the positive and negative peak values, and divides the sum by 2.
Pk Max compares the positive and negative peaks and displays the greater value (po-
larity is not indicated).
Pk+ Hold displays and holds the positive peak value until the measurement is reset.
To reset, press the MEAS RESET key, select a different detector, or re-select the same
detector.
Hold displays and holds the negative peak value until the measurement is reset.
Pk
To reset, press the MEAS RESET key, select a different detector, or re-select the same
detector.
Pk /2 Hold divides the sum of the positive and negative peak values by 2, and dis-
plays the value until the measurement is reset. To reset, press the MEAS RESET key,
select a different detector, or re-select the same detector.
Pk Mx Hold compares the positive and negative peaks and displays the greater value
until the measurement is reset. To reset, press the MEAS RESET key, select a different
detector, or re-select the same detector.
Operating Considerations
This is a priority control field. Accessing the RX TEST or TX TEST screen
overrides any changes made to this field in other screens.
See Also
"Interaction Between Screens" on page 54
Chapter 9, Duplex Test Screen
Field Descriptions
293