To Perform The Self-Calibration - HP 54645A User's And Service Manual

Mixed-signal oscilloscope
Table of Contents

Advertisement

To perform the self-calibration

In this procedure, you load the default calibration factors to give a known
starting point for the firmware calibration. However, once the default
calibration factors are loaded, you must perform the remainder of the
firmware calibration to maintain the accuracy of the oscilloscope.
Table 7-13
Equipment Required
Equipment
for HP 54645A
Pulse generator
Cable
Cable
Adapter
Adapter
Feedthrough
Equipment
for HP 54645D
Cable
Probe, 10x1
Cable
Banana Ground
Plug
Critical specifications
100 kHz, 1 V
, rise time 10 ns
p-p
BNC, 3 feet
BNC, 9 inches, Qty 2
BNC tee (m) (f) (f)
BNC (f-f)
50 , BNC (m) and (f)
Critical specifications
BNC, 3 feet
100 MHz
16-channel
Testing, Adjusting, and Troubleshooting
To perform the self-calibration
Recommended
Model/Part
HP 8112A
HP 10503A
HP 10502A
HP 1250-0781
HP 1250-0080
HP 11048C
Recommended
Model/Part
HP 10503A
HP 10074A, Qty 2
HP 54620-61801
HP 1252-7310
7-33