HP 54645A User's And Service Manual page 199

Mixed-signal oscilloscope
Table of Contents

Advertisement

Testing, Adjusting, and Troubleshooting
To construct test connectors for digital channels
Build the first test connector using a BNC connector and an 8-by-2
1
section of Berg strip.
a
b
c c
d
Figure 7-1
8-by-2 Connector Setup
Build the second test connector using a BNC connector and a 1-by-2
2
section of Berg strip.
a
b
c
Figure 7-2
1-by-2 Connector Setup
7-10
Solder a jumper wire to all pins on one side of the Berg strip.
Solder a jumper wire to all pins on the other side of the Berg strip.
Solder the center of the BNC connector to the center pin of one row
on the Berg strip.
Solder the ground tab of the BNC connector to the center pin of the
other row on the Berg strip.
Solder two resistors to the Berg strip, in parallel across the two pins.
Solder the center of the BNC connector to the center pin of one row
on the Berg strip.
Solder the ground tab of the BNC connector to the center pin of the
other row on the Berg strip.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

54645d

Table of Contents