To verify voltage measurement accuracy

To verify voltage measurement accuracy
In this test, you verify the voltage measurement accuracy by measuring the
output of a power supply using dual cursors on the oscilloscope, and
comparing the results with a multimeter.
Test limits: 1.5% of full scale 1 LSB*
Full scale is defined as 16 mV on the 1 mV/div range. Full scale on all
other ranges is defined as 8 divisions times the V/div setting.
* 1 LSB = 0.4% of full scale
Table 7-5
Equipment Required
Equipment
Power supply
Digital multimeter
Cable
Shorting cap
Adapter
Adapter
7-16
Critical specifications
14 mV to 35 Vdc, 0.1 mV resolution
Better than 0.01% accuracy
BNC, Qty 2
BNC
BNC (f) to banana (m)
BNC tee (m) (f) (f)
Recommended
Model/Part
HP 6114A
HP 34401A
HP 10503A
HP 1250-0774
HP 1251-2277
HP 1250-0781