To Perform The Self-Calibration - HP 54600B User, Installation And Service Manual

Hide thumbs Also See for 54600B:
Table of Contents

Advertisement

To perform the self-calibration

In this procedure you load the default calibration factors to give a known
starting point for the firmware calibration. However, once the default
calibration factors are loaded, you must perform the remainder of the
firmware calibration to maintain the accuracy of the oscilloscope.
Table 4-14
Equipment Required
Equipment
Pulse generator
Cable
Cable
Adapter
Adapter
Feedthrough
Check the rear panel
1
If you are not sure how to check the
output of the
Load the default calibration factors.
2
Set the rear-panel
a
(up position).
Press
b
Press the Load Defaults softkey.
c
Self-calibration hint
Thie instrument is self-calibrated at the factory. However, it should be
self-calibrated again in its working environment after a 30-minute warmup to
obtain the best accuracy.
Critical specifications
100 kHz, 1 V p-p, rise time <5 ns
BNC, 3 feet
BNC, 9 inches, Qty 2
BNC tee (m) (f) (f)
BNC (f-f)
50 Ω, BNC (m) and (f)
DC CALIBRATOR
," on page 4–6.
DC CALIBRATOR
CALIBRATION
, then press the Self Cal Menu softkey.
Print/Utility
Adjusting the Oscilloscope
Recommended
Model/Part
HP 8112A
HP 10503A
HP 10502A
HP 1250-0781
HP 1250-0080
HP 10100C
output level.
, see "To check the
DC CALIBRATOR
switch to
UNPROTECTED
Service
4–31

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 54601bHp 54602bHp 54603b

Table of Contents