To Set Up And Connect The Equipment - HP 54645A User's And Service Manual

Mixed-signal oscilloscope
Table of Contents

Advertisement

To set up and connect the equipment

Instructions for HP 54645D
These instructions apply only to the HP 54645D Mixed-Signal Oscilloscope.
Turn on the test equipment and the oscilloscope. Let them warm up
1
for 30 minutes before beginning the test.
Set up the function generator.
2
a
b
c
Using the 8-by-2 test connector and BNC cable assembly, connect
3
digital channels D0-D7 to one side of the BNC Tee. Connect the
D0-D7 ground lead to the ground side of the 8-by-2 connector.
Using a BNC-banana cable, connect the voltmeter to the other side of
4
the BNC Tee.
Connect the BNC Tee to the Main Signal output of the function
5
generator.
Figure 7-3
Threshold Test Setup
Set up the function generator to provide a DC offset voltage at the
Main Signal output.
Disable any AC voltage to the function generator output, and enable
the high-voltage output.
Monitor the function generator DC output voltage with the
multimeter.
54620A
16 CHANNEL 500 MSa/s
Testing, Adjusting, and Troubleshooting
To set up and connect the equipment
STORAGE
LOGIC ANALYZER
Measure time
Save/Recall
Entry
HORIZONTAL
TRIGGER
Delay
CHANNEL
Time/Div
Select
INPUTS
Position
Trigger out
Ext trigger in
L i n e
7-13