To Verify Bandwidth - HP 54645A User's And Service Manual

Mixed-signal oscilloscope
Table of Contents

Advertisement

To verify bandwidth

In this test you verify bandwidth by using a power meter and power sensor to
set output of a signal generator at 1 MHz and at 100 MHz. You use the
peak-to-peak voltage at 1 MHz and at 100 MHz to verify the bandwidth
response of the oscilloscope. The test is repeated at 5 mV/div (75-MHz
bandwidth).
Test limits at 10 mV/div to 5 V/div:
all channels ( 3 dB)
Test limits at <10 mV/div:
all channels ( 3 dB)
Table 7-7
Equipment Required
Equipment
Signal generator
Power meter and
Power Sensor
Power splitter
Cable
Adapter
Feedthrough
dc to 100 MHz
ac coupled 10 Hz to 100 MHz
dc to 75 MHz
ac coupled 10 Hz to 75 MHz
Critical specifications
1 to 100 MHz at 200 mV
1 to 100 MHz 3% accuracy
Outputs differ by < 0.15 dB
Type N (m), 24 inch
Type N (m) to BNC (m)
50 , BNC (m) and (f)
Testing, Adjusting, and Troubleshooting
To verify bandwidth
Recommended Model/Part
HP 8656B opt 001
HP 436A and HP 8482A
HP 11667B
HP 11500B
HP 1251-0082
HP 11048C
7-19

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

54645d

Table of Contents