Select a trace memory by pressing the
3
Press the
Save to Mem<N>
4
The current oscilloscope screen and setup are copied to the selected
trace memory.
Press the
5
Trace Mem On softkey to display memory.
The trace is displayed in half-bright video.
Automatic measurements do not operate on stored traces because stored
waveforms are pictorial information, rather than stored data.
If you have not yet changed the oscilloscope setup, and want to make
6
measurements, use the cursors.
If you changed the oscilloscope setup since you saved the display and
7
setup to trace memory, press the
cursors to make the measurements.
To clear a trace memory, press the
8
Trace Memory Operating Hint
The standard oscilloscope has volatile trace memories—saved data will be lost
when you turn the oscilloscope off. Adding an interface module, such as the
HP 54650A HP-IB Interface, or HP 54652A/B or HP 54651A RS-232 Interface,
makes the trace memory non-volatile—data is saved, even with the
oscilloscope turned off.
Saving and Recalling the Configuration
To save, recall, and clear traces
softkey.
Trace
softkey.
Recall Setup softkey. Then, use the
Clear Mem<N>
softkey.