Verifying Oscilloscope Performance

To verify bandwidth

In this test you verify the bandwidth of the oscilloscope by using a constant
amplitude signal generator. The frequency of the signal generator is set to
250 kHz to establish a reference level. Then, the frequency is changed to the
upper bandwidth limit and the level is checked to see if it is 3 dB from the
reference level.
The following procedure is a simple method to check bandwidth. However,
there is a possibility of measurement uncertainty with a constant amplitude
signal generator. If you need a more exact procedure for checking bandwidth
see, "To verify the bandwidth (alternate method)" on page 4–12.
Test limits
HP 54600B and HP 54601B, all channels (−3 dB)
HP 54602B
HP 54603B, all channels (−3 dB)
Table 4-5
Equipment Required
Equipment
Constant amplitude
signal generator
Cable
Feedthrough
4–10
dc to 100 MHz
ac coupled 10 Hz to 100 MHz.
Channels 1 & 2 (–3 dB)
10 mV/div to 5 V/div
dc to 150 MHz
1 mV/div to 5 mV/div
dc to 100 MHz
ac coupled 10 Hz to upper limit of vertical range
Channels 3 & 4 (–3 dB)
dc to 250 MHz.
dc to 60 MHz
ac coupled 10 Hz to 60 MHz.
Critical specifications
100 MHz, Constant amplitude ±1%
(250 MHz for HP 54602B)
BNC
50 Ω, BNC (m) and (f)
Recommended
Model/Part
Tek SG503/Tek TM501
HP 10503A
HP 10100C