5.4.29. Generator _Levei_Actual
Sets the generator level as in the front panel Generator "Level" function, when the
Generator Output is in "Operate" or "Operate -20dB" and the Compressor is not active.
Example of Use:
Generator _Levei_Actual
-35 . 2
- Sets the actual generator level to -35,2dB
Notes:
If Generator Output is in "Stand-by" or the Compressor is active when this message is
sent, an error will result.
If the Generator Output is in "Stand-by", the Generator_Levei_Programmed Message
can be used to set the level prior to switching On the generator output.
5.4.30. Generator _Levei_Actual?
Enables read out of the actual generator level.
Example of Use:
Generator _Levei_Actual?
enables read out of, e.g.:
G_L_A
-035 . 2dB
- If set using example in section 5.4.29
Notes:
See section 5.4.29.
5.4.31. Generator _Levei_Programmed
Sets the generator level as in the front panel Generator "Level" function, when the
Generator Output is in "Stand-by".
Example of Use:
Generator _Levei_Programmed
-47 . 5
Notes:
- Sets the programmed generator level
to -47,5dB.
The programmed generator level can be set only if the 1050 is in Generator Output
"Standby" mode. If the Generator Output "Operate" is active, an error message will
result.
If the Generator Output is in "Operate", the Generator _Levei_Actual Message can be
used to change the actual level if the compressor is not active.