Maintenance and care

Maintenance

Care and cleaning

WARNING - RISK OF ELECTRIC SHOCK
Prevent moisture from penetrating the instrument or keyboard. Disconnect
the power cable from the power supply before cleaning or disinfecting the
device.
CAUTION - SURFACE DAMAGE TO THE DEVICE
Use a disinfectant based on aldehyde and/or alcohol. The addition of
quarternary compounds is acceptable. To prevent damaging surfaces,
disinfecting agents other than those listed below must not be used
The maximum concentrations are:
• For alcohol (tested with 2-propanol): 60 %
• For aldehyde (tested with glutaraldehyde): 2 %
• For quaternary compounds (tested with DDAC): 0.2 %
CAUTION - RISK OF FALSE DIAGNOSIS
Use the supplied cover to protect the device from dust when not in use.
CAUTION - PROPERTY DAMAGE
The national disinfecting regulations must be observed in the choice of
disinfectants and disinfection procedures. Please note that some cleaning
agents and disinfectants may have an adverse effect on plastic components.
Damage caused by such disinfectants is not covered by our warranty. The
surfaces of the device have been tested and are guaranteed to resist
frequent treatment with alcoholic disinfectants and cleaning agents in the
long term.
Do not use aggressive or abrasive cleaning agents.