62
Dimensions (W x D x H)
Weight
Power supply
Power consumption
Fuses
Protection class
Ingress protection rating
Device type
Optical radiation
Battery type, manufacturer
Storage memory
Display
Chin rest
Internal printer
Type
Printing paper
Refractomy
Corneal vertex distance (VD)
Sphere(SPH)
Cylinder (CYL)
Axis (AX)
Cylinder units
Pupil distance (PD)
Minimum pupil diameter
Keratometry
Curvature radius
Corneal refraction power
Corneal astigmatism
Axis (AX)
Corneal diameter
000000-2006-367-GA-GB-151116
Technical data
Technical data
275 mm x 525 mm x 450 mm
18 kg
100 V to 240 V AC ± 10 %, 50/60 Hz
90 VA
T2AH 250 V
1
IP X0
B (DIN EN 60601-1)
Class 1 (EN ISO 15004-2)
CR2032, Hitachi Maxell
10 measurement for each eye
TFT LCD color monitor 6.4"
Electrically controlled height adjustment (max. 65 mm)
Thermal printer
Thermal paper (width 57mm, external diameter of roll 50 mm, surface
weight 60 g/m², thickness 59 µm)
0.0, 10.0, 12.0, 13.5, 15.0 mm
-25.00 D to +22.00 D (VD = 12 mm),
in steps of 0.12/0.25 D
0.00 D to ±10.00 D, in steps of 0.12 D/0. 25 D
0° to 180°, in steps of 1°
-/+/+-
10 mm to 85 mm
2 mm
5.00 mm to 10.20 mm, in steps of 0.01 mm
33.00 D to 67.50 D, in steps of 0.12 D/0.25 D
0.00 D to -15.00 D, in steps of 0.12 D/0.25 D
0° to 180°, in steps of 1°
2.0 mm to 12.0 mm, in steps of 0.1 mm