Electromagnetic compatibility

Electromagnetic compatibility
Special precautionary measures apply to this device with regard to
electromagnetic compatibility (EMC). To avoid electromagnetic disturbances,
the device may only be installed, operated and serviced in accordance with the
user manual and using the components supplied by Carl Zeiss Meditec.
CAUTION - GENERAL HAZARDS
Portable and mobile RF communications equipment may affect the device.
When operating radio devices or components for radio transmission, observe
the distances to all system components recommended in this section.
CAUTION - HAZARD DUE TO ELECTROMAGNETIC RADIATION
The VISUREF 100 may not be placed next to or stacked together with other
equipment, except in the device configurations described in this user
manual. If operation close to or together with other devices is necessary,
the VISUREF 100 must be monitored with regard to its proper functioning
in this configuration.
Replacement cables may only be purchased at Carl Zeiss Meditec.
The use of accessories, any converters or cables which are not specified in
this user manual or have not been purchased as spare parts from
Carl Zeiss Meditec can result in increased emissions or reduced immunity of
the device.