Power requirements

Power requirements

100/120 VAC, 220/240 VAC
50/60 Hz
1200 W Max (Lower range), 1300 W Max (Upper range)
The instrument can operate with mains supply voltage fluctuations up to +/-
10% of the nominal voltage.
This product does not contain customer-serviceable fuses.

Before Applying Power

Verify that all safety precautions are taken. Make all connections to the unit
before applying power. Note the external markings described under
and Symbols" on page
Getting Started Guide
"WARNING: Safety of any system incorporating the equipment is the
responsibility of the assembler of the system."
"NOTE: The input terminals for this product are classified as Measurement
Category None."
Failure to ground the test set properly can result in personal injury. Before
turning on the test set, you must connect its protective earth terminals to the
protective conductor of the main power cable. Insert the main power cable plug
into a socket outlet that has a protective earth contact only. DO NOT defeat the
earth-grounding protection by using an extension cable, power cable, or
auto-transformer without a protective ground conductor.
This instrument has auto-ranging line voltage input. Before switching on the
instrument, be sure the supply voltage is within the specified range and voltage
fluctuations do not exceed 10 percent of the nominal supply voltage..
36.
The Mains wiring and connectors shall be compatible with the connector used
in the premise electrical system. Failure to ensure adequate earth grounding by
not using the correct components may cause product damage and serious
injury.