Connection Procedures
Connecting to the IBU

Model F8820A

Figure 4-8
Connections to the IBU (Model# F8820A)
— The two IF connections (Tx and Rx) between the instruments must be
— The USB/LAN and 10 MHz Reference connections between the instruments
— Because this IBU model has only one LAN port (which is often needed for
Figure 4-9
Cable connections to the IBU (Model# F8820A)
46
duplicated for each channel in the S9165. (These connections should be
tightened with a torque wrench.
are not channel-specific; these do not need to be duplicated.
other purposes), a USB-LAN adapter is used, as shown below.
The USB connection used here is not "plug-and-play". It should be connected
before the S9165A is powered up. If it is disconnected during use, the IBU and
the S9165A will have to be rebooted after the connection is re-established.