Reading Room; Embedded Qa - Barco Nio MDNC-3521 SNES User Manual

2mp / 3mp led display
Table of Contents

Advertisement

Advanced operation
2.
Navigate to the Embedded QA/Calibration Status menu.
3.
Enter the Ambient Light Compensation submenu.
4.
Select Enabled/Disabled as desired and confirm.

4.14 Reading room

About reading room
Reading rooms can only be selected on your display when the display function is set to DICOM.
Please refer to
The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms.
Each of these reading rooms are defined by following parameters:
the maximum light allowed in this type of room
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to take
into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is enabled.
Please refer to
"Ambient Light Compensation (ALC)", page
The available reading rooms for your Nio 2MP / 3MP LED Display are:
CR/DR/ MAMMO: Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
CT/MR/NM: Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
Staff Office: Corresponds to light conditions in office rooms.
Clinical Viewing Room: Corresponds to light conditions in diagnostic reading rooms for clinical viewing.
Emergency Room: Corresponds to light conditions in emergency rooms.
Operating Room: Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
To select a reading room
1.
Bring up the OSD main menu.
2.
Navigate to the Embedded QA menu.
3.
Enter the Reading Room submenu.
4.
Select a desired reading room and confirm.

4.15 Embedded QA

About Embedded QA
Embedded QA allows you to run a display calibration or compliance test directly from the display using the
OSD menus described in the next sections. Embedded QA will use the front sensor / I-Guard to measure the
necessary luminance levels for either a calibration or compliance test. Various settings for both actions can be
selected from the display's OSD menu. The last results of both actions can be consulted from the OSD.
Embedded QA or QAWeb Enterprise?
Embedded QA is not a replacement for the Barco QAWeb Enterprise solution.
Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still
highly recommends QAWeb Enterprise as the solution of choice for calibration and QA. QAWeb Enterprise
brings many benefits such as centralized asset management, the ability to schedule tasks, remote
management, automated reporting, alerting and specific support of regional QA standards such as DIN 6868-
57, JESRA and AAPM TG18. That's why QAWeb Enterprise Agent acts as the master for all supported
32
R5915130 /05
Nio 2MP / 3MP LED Display
"Display function", page 31
31to enable ALC.