Dicom Status Report; Dicom Compliance Check; Dicom Calibration - Barco Nio MDNC-3521 SNES User Manual

2mp / 3mp led display
Table of Contents

Advertisement

displays from the moment it is installed and running. QAWeb Enterprise Agent will take over from Embedded
QA and overwrite any settings which were applied by Embedded QA.

4.15.1 DICOM status report

About DICOM status report
DICOM Compliance status: Shows if the current DICOM curve is compliant or not.
Maximum error: Shows the maximum error of the current DICOM curve. This is the deviation compared to
a perfect DICOM.
Luminance: Shows the measured luminance.
Black luminance: Shows the measured black luminance.
Time elapsed since latest compliance check: Shows the backlight runtime since last compliance check.
DICOM Calibration Status
-
Not executed: No other information is visible.
-
Calibration executed: When the calibration is executed, the time elapsed since latest calibration is
shown.
Time elapsed since latest calibration: Shows the backlight runtime since last calibration check.
DICOM compliance check: Measures the DICOM curve of your display in different steps.
DICOM calibration: Adds a correction to the current DICOM curve to approach the perfect DICOM curve
as well as possible.
DICOM Error threshold: Sets the error threshold. This is the maximum error allowed before a DICOM
calibration is required.
Ambient Light Compensation: Disables/Enables the ambient light compensation function.
Reading Room: Selects a reading room.
Reset DICOM Calibration: Restores the original (factory default) DICOM curve.
To retrieve the DICOM status report
1.
Bring up the OSD main menu.
2.
Navigate to the Embedded QA menu to make the information visible on the screen.

4.15.2 DICOM compliance check

About DICOM compliance check
The DICOM compliance check will measure the DICOM curve of your display in different steps. After
measurement, the DICOM status report is shown.
To start DICOM compliance check
1.
Bring up the OSD main menu.
2.
Navigate to the Embedded QA menu.
3.
Select DICOM Compliance Check to start the compliance check.
Warning:
Press and hold the jog dial for approximately 3 seconds during the compliance check will
abort the check.

4.15.3 DICOM calibration

About DICOM calibration
The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM curve
as well as possible.
To start DICOM calibration
1.
Bring up the OSD main menu.
2.
Navigate to the Embedded QA menu.
Advanced operation
R5915130 /05
Nio 2MP / 3MP LED Display
33