display functions have been defined. These functions emphasize on these parts containing crucial diagnostic
information by correcting the native panel behavior.
The available display functions for your Nio 2MP LED Display are:
Native: If you select Native, the native panel behavior will not be corrected.
Dynamic Gamma 1.8 or 2.2: These are gamma functions that are shifted to take into account the non-zero
luminance of an LCD panel when driven with a "black" signal. They are especially useful in CT applications
to improve the perception of low Hounsfield values.
DICOM: DICOM (Digital Imaging and Communications in Medicine) is an international standard that was
developed to improve the quality and communication of digital images in radiology. In short, the DICOM
display function results in more visible grayscales in the images. Barco recommends selecting the DICOM
display function for most medical viewing applications.
User: This display function will be automatically selected when display functions are defined by MediCal
QAWeb.
Gamma 1.8 or 2.2: Select one of these display functions in case the display is to replace a CRT display
with a gamma of 1.8 or 2.2 respectively.
The settings of the display must be adapted to suit the requirements of the visualization software. In
case of doubt, please contact the vendor of the visualization software.
To select a display function:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration menu.
3. Enter the Display Function submenu.
4. Select one of the available display functions and confirm.

4.11 Reading rooms

About reading rooms
Reading rooms can only be selected when the DICOM display function is selected. Therefore,
please refer to
"Display functions", page 24
The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms.
Each of these reading rooms are defined by following parameters:
the maximum light allowed in this type of room
The available reading rooms for your Nio 2MP LED Display are:
Dark Room: Corresponds to light conditions of ideal dark room. This setting disables the ambient light
compensation.
CR/DR/ MAMMO: Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
CT/MR/NM: Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
Staff Office: Corresponds to light conditions in office rooms.
Clinical Viewing Room: Corresponds to light conditions in diagnostic reading rooms for clinical viewing.
Emergency Room: Corresponds to light conditions in emergency rooms.
Operating Room: Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
To select a reading room:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Reading Room menu.
3. Enter the Reading Room submenu.
4. Select one of the available reading rooms and confirm.