4. Fasten the turbopump onto the counter flange in accordance with the diagram using the collar
5. Utilize all prescribed components for the turbopump.
6. Screw the hexagon head set screws into the tapped holes.
7. Tighten the hexagon head set screws in 3 stages in opposite pairs.
Connection of stud screw and tapped hole
1. Only use the approved mounting kits from Pfeiffer Vacuum for the connection.
2. Screw the required number of stud screws with the shorter screw-in end into the holes in the
3. Place the collar flange over the high vacuum flange on the turbopump.
4. Insert the snap ring into the side groove on the high vacuum flange of the turbopump.
5. Fasten the turbopump onto the counter flange in accordance with the diagram using the collar
6. Utilize all prescribed components for the turbopump.
7. Tighten the nuts in 3 stages in opposite pairs.
Connection of stud screw and through-bore
1. Only use the approved mounting kits from Pfeiffer Vacuum for the connection.
2. Place the collar flange over the high vacuum flange on the turbopump.
3. Insert the snap ring into the side groove on the high vacuum flange of the turbopump.
4. Fasten the turbopump onto the counter flange in accordance with the diagram using the collar
5. Utilize all prescribed components for the turbopump.
6. Tighten the nuts in 3 stages in opposite pairs.
7. Tightening torque: 5, 10, 16 ± 1 Nm
5.4 Connecting fore-vacuum side
Danger to life from poisoning where toxic process media leak from damaged connections
Sudden twisting of the turbopump in the event of a fault causes fittings to accelerate. There is the risk
of damaging on-site connections (e.g., fore-vacuum line) and resulting leaks. This results in leakage
of process media. In processes involving toxic media, there is a risk of injury and danger to life due to
poisoning.
► Keep masses connected to the turbopump as low as possible.
► Use flexible lines to connect to the turbopump where necessary.
flange, snap ring and centering ring.
– Observe the minimum tensile strength of the flange material and the screw-in depth.
– Tightening torque: 5, 10, 16 ± 1 Nm
counter flange.
– Observe the minimum tensile strength of the flange material and the screw-in depth.
flange, snap ring and centering ring.
– Tightening torque: 5, 10, 16 ± 1 Nm
flange, snap ring and centering ring.
WARNING
Installation
25/54