Section 3.0 System Master Interconnect Schematic (Mis Chart) - GE Optima XR642 Pre-Installation

Table of Contents

Advertisement

Section 3.0 System Master Interconnect Schematic (MIS Chart)

1
2
A
System Cabinet
(Orange)
Room
X-ray ON
Interface
Warning
Connector
Light
Door
Interlock
Switch
Shielding Cable is required
Hospital
Hospital
Vault
380-480VAC
Mains
3 Phase Supply
1
2
3
Wallbox
5336719
(No color)
Jedi CAN – A25 J108
11760A
System CAN – A25 J107
11761A
5336720
GND stud
11764A
5146500-18
120VAC
120VAC – A25 J6
11763A
5146500-17
5341237-2
RCIM2
RCIM – A25 J109
020064
TIB PWR – A25 J7
Table CAN – A25 J106
11632A
Table Ion Chamber – A25 J81
020045
020003
DPS – A25 J4
11753A
TABLE PWR – A25 J3
020005
Table RT – A25 J110
11754A
GND stud
11752A
GND stud
WS CAN – A25 J206
11644A
WS Ion Chamber – A25 J14
020021
20100A
WS DPS – A25 J2
11755A
WS PWR – A25 J1
11756A
GND stud
11757A
OTS CAN – A25 J41
11629A
P1
Stator, fan & SW – A25 J42
11710A
P2
OTS PWR – A25 J5
11708A
P3
HV Tube Anode – A9
11691A
GND stud
HV Tube Cathode – A9
11690A
GND stud
11711A
3
4
5
Hosp.
Network
5336721
11767A
5336723
11768A
20100A
5368672-4
020003
5368672-4
11776A
5336726
5336724
11770A
11774A
MONITOR
5146500-22
020065
5341236
Table (Brown)
5336731-5
J5 – Table CAN
J4 – Table Ion Chamber
5307318-5
J7- Ethernet – Table Dock
5368672-4
J2 – DPS 120VAC
5336730-5
J1 – Table 220VAC
5336729-5
J6 – Table RT bus
5336737-5
GND stud
5336736-5
Wallstand (Blue)
5146500-1
A1 J2 - Wallstand CAN
5146500-4
A1 J1 - Wallstand Ion Chamber
5368672-4
A6 J1 Ethernet – Wallstand Dock
5146500-3
J5 – Wallstand DPS 120VAC
5146500-2
J4 – Wallstand 120VAC
GND stud – A7
5146500-5
OTS (No Color)
20102A
15-pin D-Sub
OTS Console
9-pin D-Sub
20101A
OTS CAN – ALB L5
5146500-7-ROHS
Stator, fan & SW – Tube 1 Stator
5146500-9-ROHS
OTS PWR 120VAC
5146500-12-ROHS
X-ray Tube Anode
5160469-1
X-ray Tube Cathode
5160469-2
GND
5146500-8
ALB Board
REVISIONS
REV
MADE BY
DATE
PCN NO.
1
Fei xiaoyang
July 26, 2015
2
Fei xiaoyang
Aug 22, 2016
3
Li Jingliang
Apr 07, 2017
Correct J3 from 120VDC to PC RS232 to match design
4
5
6
7
System Computer (Magic)
(Yellow)
External Ethernet
USB
J2 – Jedi CAN
J1 – System CAN
USB
J4 – Ethernet – WS Dock
J5 – Ethernet – Table Dock
USB
GND stud
120VAC
USB
USB
VGA1
J6 Ethernet – Tether
USB
USB
VGA2
J3-RS232
RCIM2
Route through Cabinet
TIB (No Color)
J1-120VAC
020075
020077
Tether Ethernet
5402662
5393576-2
Tether bkp
5389155-4
Dongle
5419255
(No Color)
USB
5412201
GND stud
5411408
USB
System MIS chart without UPS
APPROVALS
DD-MM-YY
TITLE
OptimaXR646 System MIS Chart
MADE
LI Jingliang
DESCRIPTION
CHECK
First revision as DOC
FIRST MADE FOR:
Everest II
Updated
APPR
GE MEDICAL SYSTEMS
GE HUALUN MEDICAL SYSTEM
Wang Jian
MILWAUKEE, WI
OTHER
6
7
8
Control Room
A
KEYBOARD
Bar Code
Reader
MOUSE
Handswitch
C
D
7m
G
H
REV
3
DOC1484195
1
2
SHEET
OF
8

Advertisement

Table of Contents
loading

This manual is also suitable for:

Optima xr648

Table of Contents