Clear-Vu Arm Support
○
Mattress and mattress slicker
○
Shoulder Rests
○
Footrest
•
Removable rails (sleeve)
•
Rail Extender
•
Head Widener
•
Adaptor rail for table side controls
•
Width Extender
•
Armboard with thick pad (armrest)
•
Universal Clamp
•
Patient restraint strap
2.3.7 Preference Cabinet locations
NOTE:
This is applicable for all types of cabinets (including PDB).
The clear width of the service area in front of the cabinet doors to insure electrical safety shall
be at least 0.9 m for all cabinets except the Large Display cabinet (1 meter is necessary for
Large Display cabinet). In cases where 2 cabinets are installed face to face (both sides of the
access way), the clear width shall be at least 1.2 m.
Chapter 2 Equipment Requirements
Innova
IGS 520, Innova
TM
WARNING
THE HEAD HOLDER SHALL NOT BE USED TO RESTRAIN THE
PATIENT'S HEAD WHEN THE TABLE IS TILTED, ONLY WHEN IT IS
HORIZONTAL. IT IS NOT DESIGNED TO SUPPORT THE HEAD FROM
MOVING AND MAY CAUSE INJURY TO THE PATIENT.
WARNING
SLEEVE IS NOT COMPATIBLE WITH INNOVA IQ WIDE TABLETOP &
WITH HEAD EXTENDER.
WARNING
HEAD WIDENER IS NOT COMPATIBLE WITH HEAD EXTENDER &
SHOULDER REST.
IGS 530, Innova
IGS 540 Pre-Installation Manual
TM
TM
Direction 5499972-1-1EN, Revision 2
127