GE Innova IGS 520 Preinstallation Manual page 181

Table of Contents

Advertisement

Table 3-2:
Immunity Test
Electrostatic dis‐
charge (ESD)
IEC 61000-4-2
Electrical fast transi‐
ent/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply in‐
put lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Note: U
is the AC mains voltage prior to application of the test level.
n
Chapter 3 Special Construction Requirements
Innova
IGS 520, Innova
TM
IEC 60601-1-2 Test
Compliance Level
Level
+/-6 kV contact
+/-6 kV contact
+/-8 kV air
+/-8 kV air
+/-2 kV for power
+/-2 kV for power
supply lines
supply lines
+/-1 kV for input/
+/-1 kV for input/
output lines
output lines
+/-1 kV line(s) to
+/-1 kV line(s) to
lines(s)
lines(s)
+/-2 kV line(s) to
+/-2 kV line(s) to
earth
earth
0 % U
for 5 sec
0 % U
n
3 A/m
3 A/m
IGS 530, Innova
TM
Floors are wood, concrete or ceramic tile or floors are covered
with synthetic material and the relative humidity is at least 30
%.
Mains power quality is that of a typical commercial or hospital
environment.
Mains power quality is that of a typical commercial or hospital
environment.
for 5 sec
Mains power quality is that of a typical commercial or hospital
n
environment. If the user of the Innova system requires contin‐
ued operation during power mains interruptions, it is recom‐
mended that the Innova system be powered from an uninter‐
ruptible power supply.
Power frequency magnetic fields is at levels characteristic of a
typical location in a typical commercial or hospital environment.
IGS 540 Pre-Installation Manual
TM
Direction 5499972-1-1EN, Revision 2
Electromagnetic Environment
181