* Consider the 12' length of the PIMr cable, if PIMr exists.
** Pimette has 3 m cable.
NOTE:
The incorrect cable length might cause reliability issues during the operation !
Make sure, that the out of conduit parts of the cables with the connectors are properly protected
during the rest of the installation
2.4.4.2.2 For S5I GE Rev 2+ option
Use the S5I GE Rev 2+ PIM (#806365026) as reference to the procedures below.
Unpack , and check the completeness of the Preinstallation kit:
1 pcs - Custom PIM cable 27 m (88.2 ft) - Volcano ID: 806452008
1 pcs - CAT5 cable 30 m (98 ft, 5 in) - Volcano ID: 807055001
1 pcs - Custom Twin axial cable - Volcano ID: 806380001
1 pcs - Ground Cable, 30 m (98 ft, 5 in) s5i – Volcano ID: 806889001
1 pcs - S5I GE Rev 2+ Preinstallation Manual - Volcano ID: 806365007
1pcs – FFR cable 27 m - Volcano ID: 808632001
Install 806889001 Ground Cable through the Control Booth conduit (see note)
Install the cables through the S5I GE conduit from the Control Room end :
806452008 PIM cable, round shaped connector end to the Procedures Room
807055001 - Ethernet cable (1 pcs)
808632001 – FFR cable - 27 m
Install the optional cables through the S5I GE conduit from the Control Room end:
806380001 - ECG cable in case, if NON Mac-Lab ECG will be installed
NOTE:
NOTE:
NOTE:
Chapter 5 Electrical Requirements
Innova
IGS 520, Innova
TM
Cable
Control Station cable
Joystick cable
ECG cable
For the ground cable, see the S5I GE Rev 2+ service manual for special
instructions concerning cable running and cutting .
If the installation has the Mac-Lab ECG in the configuration, the ECG cable shall
not be run through the conduit, it will be used in the Control Room to connect Mac-
Lab and S5I GE PCs.
You should consider running the ECG cable through the conduit in case, the
customer requires interconnecting to a NON Mac-Lab ECG at the table base.
In this case you shall hide the unconnected cable in the conduit for future
expansion options.
IGS 530, Innova
IGS 540 Pre-Installation Manual
TM
TM
Direction 5499972-1-1EN, Revision 2
Free length from the conduit exit
2.1 m (7 ')
2.7 m (9 ')
Depends on the ECG location
229