1.1.2 Options
Table 5-2:
Option
LDM
AW
S5I GE
198
Innova
IGS 520, Innova
TM
Nominal voltage
100-120 V / 220-240 V
100-127 V / 200-240 V
100-120 V / 230 V
WARNING
PRIOR TO EACH INSTALLATION, ENSURE THAT THE ECG POWER CABLE
IS CONNECTED TO A LINE THAT IS PROTECTED AGAINST SHORT
CIRCUIT HAZARDS, ACCORDING TO LOCAL REGULATIONS AND THAT
THE ECG MONITOR IS POWERED ACCORDING TO IEC60601-1
REQUIREMENTS (LEAKAGE CURRENT AND GROUNDING).
IGS 530, Innova
TM
Frequency
Nominal Power consumption Type of power input
50 Hz or 60 Hz
3 kVA
(± 3 Hz)
50 Hz or 60 Hz
11 A / 5.5 A
50 Hz or 60 Hz
400 VA
IGS 540 Pre-Installation Manual
TM
Direction 5499972-1-1EN, Revision 2
Single phase
Single phase
Single phase
1 Power Requirements