At the end of a programmed Time Test (Timed Run or Timed Run + DD), pressing
the 2nd + GRAPH buttons will bring up a display of the Insulation Resistance
versus time curve as shown below.
The resolution of this presentation is
dependent on the sample time selected
before the test or during set-up.
to move the vertical cursor along the
time axis to display the insulation
resistance and elapsed test time
at the cursor position.
Pressing the DISPLAY button returns
to the Digital Presentation of Results.
screen to the left.
At the completion of a test, press the 2nd + TEMP buttons to show the
Temperature Correction display.
This function provides the ability to display and store temperature corrected
Insulation Resistance readings. Therefore, tests compared to historical data can
all be referenced to a common temperature for better qualitative analysis.
NOTE:
The TEMP function can be activated only after a measurement has been
completed either before or after it is stored.
If the result of your measurement is outside of the instrument's range (the
display shows the < or > symbol next to the reading), the Temperature
Correction function cannot be applied.
Temperature correction is not available on Step Voltage tests.