Procedure:
You have made a measurement
and not yet stored it. Check that
the result is within the Instrument's
measurement range.
Enter the TEMP mode by pressing
the 2nd + TEMP buttons.
Enter the estimated temperature
(Probe Temperature) at which you
made the measurement (by default,
SET-UP).
With the blinking cursor next to
button to activate probe temperature.
or decrease the temperature.
cursor back to the selection position.
Next, make sure "Resistance
Correction" is turned "on" to perform
the calculation (the default value in
"Resistance Correction." Press the
The calculation is performed immediately
and the result is displayed as Rc along
with the Reference Temperature.
This indicates what the measurement
result would have been at the
reference temperature.
The Reference Temperature (Rc
Reference Temperature) and the
T indicated and used for
SET-UP.
Note: To store this calculation, press
the 2nd + TEMP buttons again (OK is
then displayed) before storing everything.