The minimum selectable duration is 1 minute, and the maximum is
49 hours and 59 minutes.
The shortest possible sample rate is 10 seconds, and the longest
sample rate is 10:00 minutes.
again until the blinking cursor is at Timed Run.
show "OK" followed by the active test screen.
When the measurement is started, the timer counts down showing the time
remaining in the measurement.
When the Remaining Time is zero, the measurement is stopped.
During the timed run test, the intermediate samples are automatically stored.
They are used to plot a curve of insulation resistance vs time. This curve can be
displayed after the measurement is completed by pressing the Yellow 2nd button
and the GRAPH button, as long as no new measurement has been started.
resistance, if it is stored.
During the measurement, if the position of the rotary switch is changed,
or the START/STOP button is pressed, the measurement is stopped.