DEFAULT PROBE TEMPERATURE
With this function, you can program a default value for the temperature of the
device under test. This will be used when temperature correction is applied if no
value is programmed right after the test is conducted.
Rc REFERENCE TEMPERATURE
Reference temperature to which the measurement result must be referred.
Insulation resistance changes with temperature. The typical rule of thumb is that
for every 10°C increase in temperature the leakage current doubles and the
the equipment you will be testing. This value will be used when correcting results
to a reference temperature.
selected for the present test.
MAXIMUM OUTPUT VOLTAGE
The Model 5070 provides the ability to limit the maximum test voltage to a
value you specify from 40 to 5100 V. When programmed, the instrument will not
generate a higher voltage to conduct the test even if the switch position indicates
a higher voltage. For example, if you set the maximum voltage to 1250 V and
place the rotary switch in the 5000 V position and start a test, the Model 5070 will
only output 1250 V.
18
www.
Range
+5 to +167°F
(-15 to +75°C)
Range
+5 to +167°F
(-15 to +75°C)
Range
+5 to +167°F
-15 to +75°C
Range
40 to 5100 V
Megohmmeter Model 5070