AEMC 5070 User Manual page 49

Megohmmeters
Hide thumbs Also See for 5070:
Table of Contents

Advertisement

1.800.561.8187
You select an insulation measurement
with one of the selectable test voltages.
Switch Position: 50 to 5000 V
You can choose from one of the three
another voltage by selecting it with the
buttons. The new selected voltage will
not be saved when the instrument is
another position.
You select an insulation measurement
with a test voltage that varies in steps
(this is the "Step Function" mode).
Switch Position:
You can choose from the three stored
If the beeper is selected as "ON" in the SET-UP mode, an audible beep is emitted
every 10 seconds to indicate that a test is in process.
Some special functions can be performed during the test and are described in
Chapter 3.
WARNING: Insulation measurements cannot be started if there is an
excessively high external voltage on the "+" and "–" terminals.
The Model 5070 will automatically inhibit testing.
and described next, the insulation test is not started and an audible alarm is
Megohmmeter Model 5070
www.
, if the external voltage of the terminals
information@itm.com
.com
49