SET DEFAULT PARAMETER
This function allows you to reset all the user programmable functions back to the
factory defaults.
CLEAR MEMORY
The clear memory function allows you to selectively erase individual tests or all
ensure that you don't do this by accident.
the selected measurement voltage. If the V Disturbance exceeds the value of
the ratio, the instrument stops the test in process. If the voltage exceeding the
ratio is present at the device under test before the test is started, a test will not
be allowed to start. Three programmable values are: 3%, 10% and 20% of the
measurement voltage. The default is 3%.
If a test is to be performed at 1000 V and the V Disturbance / V Output
is set at 10%, the presence of 100 volts (1000 V * 10% = 100) before the test
starts will inhibit the test.
BUZZER
The Model 5070 is equipped with a buzzer that will emit an audible tone when a
key is pressed, at regular intervals during a timed test, or continuously during an
POWER DOWN
5 minutes of no activity, if a timed test is not in progress. This function lets you
Megohmmeter Model 5070
www.
Range
3%, 10%, or 20%
Range
ON or OFF
Range
ON or OFF