Finder mode:
In the Finder mode, the sound frequency depends on the pulse
rate of the detector. The more detector pulses above the count
rate that is present at the moment of the activation of the finder
mode are captured by the detector, the higher the tone.
The finder mode must be enabled in the menu.
It is activated and deactivated by two times pressing the right or
top button:
4.2

Alarm indication

Each time the first alarm threshold is exceeded, the alarm de-
vices beeper, LED and the vibrator become active, if they are
enabled.
Alarm 1: LED slowly blinking, two frequency alarm tone
Alarm 2: LED quick blinking, continuous alarm tone
Dose Alarm: LED constantly on, continuous alarm tone, vi-
brator slow.
4-2
DB-059-060324 E
Iw/Pi/Ff 30.07.2008
Indication of activated
finder mode
SENTINEL RadEye G