Portable dose- and dose rate meter
Table of Contents

Advertisement

5-2
DB-059-060324 E
Iw/Pi/Ff 30.07.2008
Error high voltage generation
No detector pulse within 128 seconds
EEPROM with calibration data shows
EEPROM Read or EEPROM Write
error.
A Watchdog Error indicates, that the
micro controller has problems to work
on it's tasks in a given timeframe.
Reasons are strong electromagnetic
pulses, firmware errors or hardware
issues.