8.3
Firmware revisions
V1.12 S
- Detector supervision time out set for 120 s. Display shows
0.00 mR/h until the first pulse. After the first detector pulse
the measurement is started.
V1.19 S
- Battery voltage supervision adapted to RadEye car holder.
- Menu functions in German and French.
- New menu function "Show alarm" added
- New menu function "Backlight" added
- New menu function "Settings" for type of battery added. It is
used for "Batt low" warning.
- Date format according to ISO 8601:
JJ-MM-DD hh:mm
- New function "Alarm latching" added.
- Error in previous version:
dose measurement only to 68 R. Then overrun to 0 R. No
overload display V1.19 S dose measurement to 1000 R.
The overload display.
- Background subtraction for dose and dose rate unified to
0.0625 cps.
SENTINEL RadEye G
Iw/Pi/Ff 30.07.08
DB-059-060324 E
8-7