Metering Keys

Metering keys control the metering functions of the dc source. As set from the factory, all front panel
measurements are calculated from a total of 2048 readings taken at a 46.8 microsecond sampling rate.
Therefore, the factory default acquisition time for a single front panel measurement is about 100
milliseconds. Refer to "Making Front Panel Measurements" for more information about changing the
front panel sampling rate and the number of measurement points.
NOTE:
The front panel sample rate and data point settings are separate and independent of the
sample rate and data point settings that are programmed over the HP-IB interface. When
an HP-IB measurement is in progress, the front panel display temporarily indicates
-- -- -- -- --. Front panel measurements resume when the HP-IB measurement completes.
HP-IB measurements are discussed in chapter 7.
Press this key to access the meter menu list. Also use this key to exit a menu at
Meter
any time and return to meter mode.
Display
<reading>V <reading>A
<reading>V MAX
<reading>V MIN
<reading>V HIGH
<reading>V LOW
<reading>V RMS
<reading>A MAX
<reading>A MIN
<reading>A HIGH
<reading>A LOW
<reading>A RMS
Press this key to access the following metering functions.
Shift
Input
Display
CURR:RANGE <char>
CURR:DET <char>
TINT <value>
POINTS <char>
Notes:
reading = the returned measurement
value = a numeric value
char = a character string parameter
Use
Use
Use
Measurement
Measures output dc voltage and current
Measures peak output voltage
Measures minimum output voltage
Measures the high level of a voltage pulse waveform
Measures the low level of a voltage pulse waveform
Measures rms voltage
Measures peak output current
Measures minimum output current
Measures the high level of a current pulse waveform
Measures the low level of a current pulse waveform
Measures rms current
Command Function
Select current range (AUTO, LOW or HIGH)
Select current measurement bandwidth (ACDC or DC)
Sets the front panel measurement interval in seconds
(15.6 µs to 1 second)
Sets the number of points in the front panel measurement
buffer ( 1, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, 2048)
and
to scroll through the menu commands.
and
to scroll through the menu parameters.
and
to select a digit in a numeric entry field.
Front Panel Operation - 5
37