Trigger:sequence2:Slope:current Trigger:acquire:slope:current; Trigger:sequence2:Slope:voltage Trigger:acquire:slope:voltage - Agilent Technologies 66311A User Manual

Table of Contents

Advertisement

TRIGger:SEQuence2:SLOPe:CURRent
TRIGger:ACQuire:SLOPe:CURRent
This command sets the slope of an internally triggered current measurement.
triggering occurs on the rising edge.
POSitive
NEGative
triggering occurs on the falling edge.
EITHer
triggering occurs on either edge.
Command Syntax TRIGger:SEQuence2:SLOPe:CURRent<slope>
Parameters EITHer | POSitive | NEGative
*RST Value POSitive
Examples TRIG:SEQ2:SLOP:CURR POS
Query Syntax TRIGger:SEQuence2:SLOPe:CURRent?
Returned Parameters <CRD>
Related Commands TRIG:SEQ2:SLOP:VOLT
TRIGger:SEQuence2:SLOPe:VOLTage
TRIGger:ACQuire:SLOPe:VOLTage
This command sets the slope of an internally triggered voltage measurement.
triggering occurs on the rising edge.
POSitive
NEGative
triggering occurs on the falling edge.
triggering occurs on either edge.
EITHer
Command Syntax TRIGger:SEQuence2:SLOPe:VOLTage<slope>
Parameters EITHer | POSitive | NEGative
*RST Value POSitive
Examples TRIG:SEQ2:SLOP:VOLT POS
Query Syntax TRIGger:SEQuence2:SLOPe:VOLTage?
Returned Parameters <CRD>
Related Commands TRIG:SEQ2:SLOP:CURR
TRIGger:ACQuire:SLOPe:CURRent<slope>
TRIGger:ACQuire:SLOPe:CURRent?
TRIGger:ACQuire:SLOPe:VOLTage<slope>
TRIGger:ACQuire:SLOPe:VOLTage?
Language Dictionary - 8
TRIG:ACQ:SLOP:CURR EITH
TRIG:ACQ:SLOP:VOLT EITH
111

Advertisement

Table of Contents
loading

Table of Contents