TRIGger

When the transient trigger subsystem is initiated, this command generates a trigger signal. The trigger
will then:
1. Initiate a pending level change as specified by CURRent:TRIGger or VOLTage;TRIGger.
2. Clear the WTG bit in the Status Operation Condition register after both transient and acquire trigger
sequences have completed. (WTG is the logical-or of both transient and acquire sequences.)
3. If INITiate:CONTinuous ON has been programmed, the trigger subsystem is immediately re-enabled
for subsequent triggers. As soon as it is cleared, the WTG bit is again set to 1.
Command Syntax TRIGger[:SEQuence1][:IMMediate]
Parameters None
Examples TRIG
Related Commands ABOR

TRIGger:SOURce

This command is included for completeness. It selects the trigger source for transient triggers. Since BUS
is the only trigger source for transient triggers, this command does not need to be used.
HP-IB device, *TRG, or <GET> (Group Execute Trigger)
BUS
Command Syntax TRIGger[:SEQuence1]:SOURce<source>
Parameters BUS
*RST Value BUS
Examples TRIG:SOUR BUS
Query Syntax TRIGger[:SEQuence1]:SOURce?
Returned Parameters <CRD>
TRIGger:SEQuence2
TRIGger:ACQuire
When the trigger subsystem is initiated, these commands generate a measurement trigger signal. The
measurement trigger causes the dc source to measure the output voltage and current and store the results
in a buffer.
Command Syntax TRIGger:SEQuence2[:IMMediate]
Parameters None
Examples TRIG:SEQ2
Related Commands TRIG:SOUR
TRIGger[:TRANsient][:IMMediate]
TRIG:IMM
CURR:TRIG
TRIGger[:TRANsient]:SOURce<source>
TRIGger[:TRANsient]:SOURce?
.
TRIGger:ACQuire:[:IMMediate]
TRIG:SEQ2:DEF
TRIG:SEQ2:LEV:VOLT
Language Dictionary - 8
INIT
*TRG
VOLT:TRIG
TRIG:ACQ
TRIG:SEQ2:COUN
TRIG:SEQ2:SLOP:CURR
107