MEASure:CURRent?
FETCh:CURRent?
These queries return the dc output current.
Query Syntax MEASure[:SCALar]:CURRent[:DC]?
Parameters None
Examples MEAS:CURR? MEAS:CURR:DC?
Returned Parameters <NR3>
Related Commands MEAS:VOLT?
MEASure:CURRent:ACDC?
FETCh:CURRent:ACDC?
These queries return the ac+dc rms output current.
Query Syntax MEASure[:SCALar]:CURRent:ACDC?
Parameters None
Examples MEAS:CURR:ACDC?
Returned Parameters <NR3>
Related Commands MEAS:VOLT:ACDC?
MEASure:CURRent:HIGH?
FETCh:CURRent:HIGH?
These queries return the High level current of a current pulse waveform. The instrument first measures
the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse
waveform using 1024 bins between the maximum and minimum data points. The bin containing the most
data points above the 50% point is the high bin. The average of all the data points in the high bin is
returned as the High level. If no high bin contains more than 1.25% of the total number of acquired
points, then the maximum value is returned by these queries.
Query Syntax MEASure[:SCALar]:CURRent:HIGH?
Parameters None
Examples MEAS:CURR:HIGH?
Returned Parameters <NR3>
Related Commands MEAS:CURR:LOW?
FETCh[:SCALar]:CURRent[:DC]?
FETCh[:SCALar]:CURRent:ACDC?
FETCh[:SCALar]:CURRent:HIGH?
Language Dictionary - 8
FETC:CURR:ACDC?
FETC:CURR:HIGH?
CALC:REF:HIGH
87