CALibrate:DATA

This command enters a calibration value that you obtain by reading an external meter. You must first
select a calibration level (with CALibrate:LEVel) for the value being entered.
Command Syntax CALibrate:DATA<NRf>
Parameters <external reading>
Examples CAL:DATA 3222.3 MA
Related Commands CAL:STAT

CALibrate:LEVel

This command selects the next point in the calibration sequence.
P1: the first calibration point
P2: the second calibration point
Command Syntax CALibrate:LEVel <point>
Parameters P1 | P2
Examples CAL:LEV P2

CALibrate:PASSword

This command lets you change the calibration password. A new password is automatically stored in
nonvolatile memory and does not have to be stored with CALibrate:SAVE.
If the password is set to 0, password protection is removed and the ability to enter the calibration mode is
unrestricted.
Command Syntax CALibrate:PASScode<NRf>
Parameters <model number> (default)
Examples CAL:PASS 6812
Related Commands CAL:SAV

CALibrate:SAVE

This command saves any new calibration constants after a calibration procedure has been completed in
nonvolatile memory. If CALibrate:STATe OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored..
Command Syntax CALibrate:SAVE
Parameters None
Examples CAL:SAVE
Related Commands CAL:PASS
Unit A (amperes)
CAL:LEV
CAL:STAT
Language Dictionary - 8
CAL:DATA 5.000
CAL:PASS 6.1994
83