8 – Language Dictionary

CALibrate:STATe

This command enables and disables calibration mode. The calibration mode must be enabled before the
dc source will accept any other calibration commands.
The first parameter specifies the enabled or disabled state. The second parameter is the password. It is
required if the calibration mode is being enabled and the existing password is not 0. If the password is
not entered or is incorrect, an error is generated and the calibration mode remains disabled. The query
statement returns only the state, not the password.
NOTE:
Whenever the calibration state is changed from enabled to disabled, any new calibration
constants are lost unless they have been stored with CALibrate:SAVE.
Command Syntax CALibrate:STATe<bool>[,<NRf>]
Parameters 0 | 1 | OFF | ON [,<password>]
*RST Value OFF
Examples CAL:STAT 1,6812
Query Syntax CALibrate:STATe?
Returned Parameters <NR1>
Related Commands CAL:PASS

CALibrate:VOLTage

This command initiates the calibration of the output voltage and the voltage readback circuit.
Command Syntax CALibrate:VOLTage[:DC]
Parameters None
Examples CAL:VOLT

CALibrate:VOLTage:PROTection

This command calibrates the overvoltage protection (OV) circuit. The dc source automatically performs
the calibration. CALibrate:VOLTage:PROTection is a sequential command that takes several seconds to
complete.
Command Syntax CALibrate:VOLTage:PROTection
Parameters None
Examples CAL:VOLT:PROT
84
CAL:STAT OFF
CAL:SAVE
*RST
CAL:VOLT:DC