Trigger:sequence2:Count:current Trigger:acquire:count:current - Agilent Technologies 66311A User Manual

Table of Contents

Advertisement

8 – Language Dictionary
TRIGger:SEQuence2:COUNt:CURRent
TRIGger:ACQuire:COUNt:CURRent
This command sets up a successive number of triggers for measuring current data. With this command,
the trigger system needs to be initialized only once at the start of the acquisition period. After each
completed measurement, the instrument waits for the next valid trigger condition to start another
measurement. This continues until the count has completed.
Command Syntax TRIGger:SEQuence2:COUNt:CURRent<NRf+>
Parameters 1 to 100
*RST Value 1
Examples TRIG:SEQ2:COUN:CURR 5
Query Syntax TRIGger:SEQuence2:COUNt:CURRent?
Returned Parameters <NR3>
Related Commands TRIG:SEQ2
TRIGger:SEQuence2:COUNt:VOLTage
TRIGger:ACQuire:COUNt:VOLTage
This command sets up a successive number of triggers for measuring voltage data. With this command,
the trigger system needs to be initialized only once at the start of the acquisition period. After each
completed measurement, the instrument waits for the next valid trigger condition to start another
measurement. This continues until the count has completed.
Command Syntax TRIGger:SEQuence2:COUNt:VOLTage<NRf+>
Parameters 1 to 100
*RST Value 1
Examples TRIG:SEQ2:COUN:VOLT 5
Query Syntax TRIGger:SEQuence2:COUNt:VOLTage?
Returned Parameters <NR3>
Related Commands TRIG:SEQ2
108
TRIGger:ACQuire:COUNt:CURRent<NRf+>
TRIGger:ACQuire:COUNt:CURRent?
TRIG:ACQ
TRIGger:ACQuire:COUNt:VOLTage<NRf+>
TRIG:ACQ:COUN:VOLT 1
TRIGger:ACQuire:COUNt:VOLTage?
TRIG:ACQ
TRIG:ACQ:COUN:CURR 1