Current Ranges And Measurement Detector; Window Functions - Agilent Technologies 66311A User Manual

Table of Contents

Advertisement

7 - Programming the DC Source
When the instrument is turned on and at *RST, the output voltage or current sampling rate is 15.6
microseconds, and the sweep size is set to 2048 data points. This means that it takes about 32
milliseconds to fill up 2048 data points in the data buffer. Adding a command processing overhead of
about 20 milliseconds results in a total measurement time of about 50 milliseconds per measurement.
You can vary this data sampling rate with:
SENSe:SWEep:TINTerval <sample_period>
SENSe:SWEep:POINts <points>
For example, to set the time interval to 46.8 microseconds per measurement with 1500 samples, use
SENSe:SWEep:TINTerval 46.8E-6;POINts 1500.
Note that reducing the number of sample points increases the speed of the measurement; however, the
tradeoff is greater measureent uncertainty in the presence of noise..

Current Ranges and Measurement Detector

The dc source has two current measurement ranges. The command that controls the ranges is:
SENSe:CURRent:RANGe MIN | MAX
When the range is set to MIN, the maximum current that can be measured is 20 milliamperes. The
crossover value of the high and low ranges is 20 milliamperes.
The dc source also has two measurement detectors. Check that the current detector is set to ACDC when
measuring current pulses or other waveforms with a frequency content greater than a few kilohertz.
SENSe:CURRent:DETect ACDC
Select DC as the measurement detector if you are making only DC current measurements and you require
a measurement offset better than 2mA on the High current measurement range. Note that this selection
gives inaccurate results on current waveforms that have ac content.
SENSe:CURRent:DETect DC

Window Functions

The dc source lets you select from two measurement window functions: Hanning and Rectangular. To
select a window function, use:
SENSe:WINDow: HANN | RECT
As shipped from the factory, the dc source measurement functions use a Hanning window. The Hanning
4
window applies a cos
weighting function to the data in the measurement buffer when computing
measurements such as average and rms. This returns accurate data even if an integral number of
waveform cycles are not captured, provided that at least three or more waveform cycles are in the
measurement buffer. If there are only one or two waveform cycles, the Hanning window will not give
accurate results.
With a Rectangular window, no weighting function is applied to the data in the measurement buffer.
However, to use the Rectangular window function to return accurate data for one or more waveform
cycles, an integral number of of waveform cycles must be captured in the measurement buffer. This
means that you must accurately know the waveform period beforehand. In this way you can chose the
62

Advertisement

Table of Contents
loading

Table of Contents