Agilent Technologies 66311A User Manual page 154

Table of Contents

Advertisement

F – Compatibility Language
Compatibility
Command
UNMASK
These commands determine the conditions that will set bits in the
xxx
fault register, allowing the operator to define the conditions that
will be reported as fault Fault conditions can be enabled by
sending the decimal equivalent of the total bit weight of all
conditions to be enabled.
VOUT?
This command measures and returns the actual output voltage.
Data Representation: SZZD.DD; (SZD.DDD for 6634B only)
VSET <n>
This command programs the output voltage. See Table 8-3 for the
programming range of this command.
Initial condition: 0 V
A = Alpha
D = Digit
S = Sign (blank for positive and – for negative)
Z = Digit with leading zeros output as spaces
Error
ERR 0
No error
ERR 1
EEPROM save failed [Data write to non-volatile memory failed]
ERR 2
Second PON after power-on [More than one PON command received after power-on. Only
one is allowed.]
ERR 4
RLYPON sent with no relay option present [A RLYPON command was sent with no relay
option present.]
ERR 5
No relay option present [A relay option command was sent with no relay option present.]
ERR 8
Addressed to talk and nothing to say [The unit was addressed to talk without first receiving a
query.]
ERR 10
Header expected [A non-alpha character was received when a header was expected.]
ERR 11
Unrecognized header [The string of alpha characters received was not a valid command.]
ERR 20
Number expected [A non-numeric character was received when a number was expected.]
ERR 21
Number Syntax [The numeric character received did not represent a proper number.]
ERR 22
Number out of internal range [The number received was too large or small to be represented in
internal format.]
ERR 30
Comma [A comma was not received where one was expected.]
ERR 31
Terminator expected [A valid terminator was not received where one was expected.]
ERR 41
Parameter Out [The number received exceeded the limits for its associated command.]
ERR 42
Voltage Programming Error [The programmed value exceeded the valid voltage limits.]
ERR 43
Current Programming Error [The programmed value exceeded the valid current limits.]
ERR 44
Overvoltage Programming Error [The programmed value exceeded the valid overvoltage
limits.]
ERR 45
Delay Programming Error [The programmed value exceeded the valid delay limits.]
ERR 46
Mask Programming Error [The programmed value exceeded the fault mask limits.]
ERR 51
EEPROM Checksum [EEPROM failed, or a new uncalibrated EEPROM was installed.]
152
Table F-2. COMPatibility Commands (continued)
Description
Table F-3. COMPatibility Errors
NumberError String [Description/Explanation/Examples]
Similar SCPI
Command
STAT:OPER:ENAB
STAT:QUES:ENAB
*ESE
MEAS:VOLT?
VOLT

Advertisement

Table of Contents
loading

Table of Contents