*TRG

This common command generates a trigger when the trigger subsystem has BUS selected as its source.
The command has the same affect as the Group Execute Trigger (<GET>) command.
In RS-232 mode, this command emulates some of the functionality of the IEEE-488 Group Execute
Trigger command.
Command Syntax *TRG
Parameters None
Related Commands ABOR INIT TRIG[:IMM]

*TST?

This query causes the dc source to do a self-test and report any errors. 0 indicates that the dc source
passed self-test. 1 indicates that one or more tests failed. Selftest errors are written to the error queue (see
Appendix C).
Query Syntax TST?
Returned Parameters <NR1>

*WAI

This command instructs the dc source not to process any further commands until all pending operations
are completed. "Pending operations" are as defined under the *OPC command. *WAI can be aborted
only by sending the dc source an HP-IB DCL (Device Clear) command.
Command Syntax WAI?
Parameters None
Related Commands *OPC*OPC?